Method development validation for corticoids in animal feed samples by liquid chromatography using a monolithic column

Thumbnail
View/Open
IIT-07-059A.pdf (803.9Kb)
Date
01/11/2007
Author
Muñíz Valencia, Roberto
Gonzalo Lumbreras, Raquel
Santos Montes, Ana María
Izquierdo Hornillos, Roberto
Estado
info:eu-repo/semantics/publishedVersion
Metadata
Show full item record
Mostrar METS del ítem
Ver registro en CKH

Refworks Export

Abstract
 
 
A LC method for corticosteroids (CC) determination in poultry feed using a Chromolith column and UV detection has been developed and validated. The method development involved the optimization of different hydro-organic mobile phases using methanol or ACN as organic modifiers, flow rate, and temperature. The optimum separation was achieved at 40 degrees C using ACN/water (21:79 v/v) as mobile phase and 3 mL/min flow rate, allowing the separation to baseline of four out of seven CC in about 10 min. Prior to LC, a sample preparation procedure previously assayed for anabolics was used. It includes a leaching process, saponification of the esters from fatty acids, and SPE. Method validation was carried out according to the EU criteria established for quantitative screening methods. The extraction efficiencies, decision limits (CCalpha), and detection capabilities (CCbeta) for these compounds were in the ranges of 86-92\%, 27-36 microg/kg, and 33-43 microg/kg, respectively. The repeatability and the within-laboratory reproducibility at 1, 1.5, and 2 CCbeta concentration levels were smaller than 9.0, 5.0, and 4.2\% and 9.4, 6.4, and 4.9\%, respectively. The CV values of the robustness test were less than 3.8\% and the accuracy was in the range of 98-103\%. The proposed method was applied to other feed with satisfactory results.
 
URI
http://hdl.handle.net/11531/5266
Method development validation for corticoids in animal feed samples by liquid chromatography using a monolithic column
Tipo de Actividad
Artículos en revistas
ISSN
1615-9306
Palabras Clave

Animal feed; Corticoids; Monolithic column
Collections
  • Artículos

Repositorio de la Universidad Pontificia Comillas copyright © 2015  Desarrollado con DSpace Software
Contact Us | Send Feedback
 

 

Búsqueda semántica (CKH Explorer)


Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsxmlui.ArtifactBrowser.Navigation.browse_advisorxmlui.ArtifactBrowser.Navigation.browse_typeThis CollectionBy Issue DateAuthorsTitlesSubjectsxmlui.ArtifactBrowser.Navigation.browse_advisorxmlui.ArtifactBrowser.Navigation.browse_type

My Account

LoginRegister

Repositorio de la Universidad Pontificia Comillas copyright © 2015  Desarrollado con DSpace Software
Contact Us | Send Feedback