Please use this identifier to cite or link to this item: http://hdl.handle.net/11531/5396
Title: Determination of carbendazim in soil samples by anodic stripping voltammetry using a carbon fiber microelectrode
Authors: González de la Huebra, Mª José
Hernández Hernández, Pedro
Nieto Palmeiro, Oscar
Ballesteros Iglesias, María Yolanda
Hernández Hernández, Lucas
Issue Date:  1
Abstract: 
A method for the determination of carbendazim (MBC) by anodic stripping voltammetry using a carbon fiber ultramicroelectrode was developed. The ultramicroelectrode was made in our laboratory and its electrochemical behavior was characterized by measuring the electrochemical response with a solution of potassium ferricyanide. The optimum parameters used for the determination of MBC are the following: 0.05 M phosphate buffer at pH 2.0 as supporting electrolyte; a scan rate of v = 10.00 V s 1 and an accumulation potential of Eac = 0.00 V. The MBC was determined in a soil sample with the method proposed and the results found were comparable to those obtained by HPLC.
Description: Artículos en revistas
URI: http://hdl.handle.net/11531/5396
ISSN: 1618-2642
Appears in Collections:Artículos

Files in This Item:
File Description SizeFormat 
IIT-00-102A.pdf432,91 kBAdobe PDFView/Open    Request a copy


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.