Por favor, use este identificador para citar o enlazar este ítem: http://hdl.handle.net/11531/87064
Título : Short communication: Torsional tests on long-bone cylindrical medullary-canal exo-endoprostheses
Autor : Mendaza De Cal, Rosa María
Ballesteros Iglesias, María Yolanda
Peso Fernández, Salvador
del Real Romero, Juan Carlos
Rodríguez Quirós, Jesús
Fecha de publicación : 1-jun-2023
Resumen : Torsion tests are performed for cylindrical type of a novel exo-endoprosthesis internal part. It is verified that it has values greater than those obtained for elliptical type. The best bonding method between PEEK and threaded rod is the resin adhesive, matching the obtained results for elliptical implant. Results of cylindrical type are discussed versus elliptical type.
Torsion tests are performed for cylindrical type of a novel exo-endoprosthesis internal part. It is verified that it has values greater than those obtained for elliptical type. The best bonding method between PEEK and threaded rod is the resin adhesive, matching the obtained results for elliptical implant. Results of cylindrical type are discussed versus elliptical type.